Category: Volume
Part: 4-STP
Section: 21
Publication date:
Revision number: 1
Part 4, section 21—Specialized test procedure—Flow rate determination
Application
This procedure is used when performing metering system accuracy tests.
Purpose
This procedure is used to determine the approximate flow rate at which a meter is tested and to establish whether the flow rate exceeds the meter/system capacity.
Reference
R.271, R.280, R.290 of the Weights and Measures Regulations and Notice of Approval.
Procedure
Maximum flow rates are determined with the nozzle or loading valve fully opened. Minimum flow rates are determined in order to maintain the test run at or slightly above the minimum approved flow rate for the meter. This procedure is applied during the fast and slow tests to determine whether the operating flow rates are within the approved limits of the meter.
Stabilize the flow rate for the test run (e.g., run approximately 100 L of a 500 L test run).
While maintaining the flow rate (i.e., do not stop or change the flow rate), time an even number of litres or kilograms with a stopwatch.
The flow rate is calculated as follows:
Or, for Coriolis effect meters measuring in mass units:
Examples
If it takes 15 seconds to run 100 L, the approximate flow rate is:
If it takes 45 seconds to run 90 kg of product, the approximate flow rate is:
Interpretation of results
If the system is capable of operating beyond the approved maximum flow rate of the meter, an automatic flow control valve must be installed downstream from the meter.
For accuracy tests, the meter must run at flow rates between its approved minimum and maximum.
Revisions
The purpose of revision 1 was to add the procedure for determining kilograms per minute for mass to mass testing of Coriolis effect meters.